IEC 60601-1-Figure 34 Spark Ignition Test Apparatus

অন্যান্য ভিডিও
November 28, 2022
সংক্ষিপ্ত: Discover the IEC 60601-1-Figure 34 Spark Ignition Test Apparatus, designed to assess fire risks in oxygen-rich environments for medical equipment. This essential tool ensures compliance with safety standards, reducing ignition risks under normal and fault conditions.
সম্পর্কিত পণ্য বৈশিষ্ট্য:
  • Evaluates fire risks in oxygen-rich environments for medical equipment.
  • Tests ignition sources under normal and single fault conditions.
  • Includes a spark ignition test apparatus with specific contact pin configurations.
  • Determines safe upper limits for voltage and current with a safety margin factor.
  • Conducts over 300 trials to ensure no ignition occurs.
  • Complies with IEC 60601-1 standards for medical electrical equipment.
  • Features detailed test parameters including oxygen concentration and fuel type.
  • Ensures safety by limiting fire spread in high-risk environments.
FAQS:
  • What is the purpose of the IEC 60601-1-Figure 34 Spark Ignition Test Apparatus?
    The apparatus is used to assess and reduce fire risks in oxygen-rich environments for medical equipment, ensuring compliance with safety standards under normal and fault conditions.
  • How many trials are conducted during the spark ignition test?
    A minimum of 300 trials are performed to determine if sparks do not ignite, ensuring reliable results under worst-case scenarios.
  • What standards does the Spark Ignition Test Apparatus comply with?
    The apparatus complies with IEC 60601-1 standards, which outline general requirements for basic safety and essential performance of medical electrical equipment.